Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
9-32
Install Solenoid Valves
1. Disassemble replacement solenoid valve (15 or 16, Figure 9-24).
2. Install new solenoid valve base. Be careful not to overtighten.
3. Install new washer and solenoid coil assembly and secure with nut.
4. Connect the solenoid leads to the proper terminations on the solenoid
power terminal block (Figure 9-27).
Install Electronics Stack
1. Install electronics stack (5, Figure 9-24) and secure with lockwashers
(14) and screws (13).
2. See Figure 9-26 and Figure 9-27. Reconnect power cable, signal cable,
and solenoid lead connectors to electronics stack terminals.
Install LOI Module and Board
1. Install LOI board (9, Figure 9-24) and secure with two screws (11) and
lockwashers (12).
2. Note the location of the LOI connector (10). Plug LOI module (8) and
connector into one of the four mating receptacles provided.
3. Install three screws (7) to secure the LOI module.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...