Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
7-15
Parameter Mnemonic
Valid Range
Units
Description
COZERO_PURGE_TIME
60-180
Seconds
The duration of the Combustibles sensor zero purge.
COZERO_STATE
0, 1, 2
The current step of the Combustibles sensor zero cycle. (0=Idle,
1=Flowing, 2=Purging)
COZERO_UPDATE
0: Off
– 1: On
Enumerated
Indicates whether the Combustibles calibration constants should be
updated after Combustibles sensor zero.
DETAILED_STATUS_1
0-16777215
Bit Enum
A bit-enumerated value used to communicate the status of the OCX. (This
is similar in nature to the command 48 status bits in HART).
DETAILED_STATUS_2
0-16777215
Bit Enum
A bit-enumerated value used to communicate the status of the OCX. (This
is similar in nature to the command 48 status bits in HART).
ELECTRONICS_TEMP
°C
The current temperature reading of the electronics temperature sensor.
ELECTRONICS_TEMP_
mV
The current voltage reading of the electronics temperature sensor.
INPUT
INITIATE_BLOWBACK
1=Do a manual
This initiates a blowback cycle. 1=Do a manual blow back
blowback
LINE_FREQUENCY
Hz
The Calculated line frequency.
LINE_VOLTAGE
Volts
The calculated line voltage.
MAX_ELECTRONICS_TEMP
°C
This is the maximum electronics temperature seen by the analyzer.
MAX_TEMP_RESET
1=Reset max
This resets the maximum temperatures.
temperatures
MODE_BLK
See FF-891 section 5.3.
O2_AUTOCAL_INTERVAL
0-9999
Hours
The time between automatic calibrations of the O
2
sensor.
O2_CAL_POINT_HI
0-40
%
The value of the O
2
high test gas. This gas is also used as the low gas for
calibrating the combustibles sensor.
O2_CAL_POINT_LO
0-40
%
The value of the O
2
low test gas.
O2_CELL_IMPEDANCE
Ohms
The instaneous impedance value for the O
2
cell.
O2_CONSTANT
-20.0-20.0
mV
The O
2
calibration constant.
O2_FAILED_CONSTANT
mV
This is the constant value calculated from the last failed O
2
calibration.
O2_FAILED_SLOPE
mV/Decade
This is the slope value calculated from the last failed O
2
calibration.
O2_HTR_DUTYCYCLE
O
2
heater duty cycle.
O2_IMPEDANCE CAL
Ohms
The impedance value that was calculated as a result of the current
successful O
2
calibration.
O2_PERCENT_OF_RANGE
%
The percent of range of the current O
2
reading.
O2_PREVIOUS_CONSTANT
mV
The O
2
calibration constant from the previous good calibration.
O2_PREVIOUS_
Ohms
The impedance value from the previous good calibration.
IMPEDANCE
O2_PREVIOUS_SLOPE
mV/Decade
The O
2
calibration slope from the previous good calibration.
O2_PRIMARY_VALUE
The value and status of the O
2
concentration reading.
O2_PRIMARY_VALUE_
The High and Low range limit values, the engineering units code, and the
RANGE
number of digits to the right of the decimal point for the O
2
reading.
O2_PRIMARY_VLUE_TYPE See F-903
See FF-903 section 3.3.
section 4.1
O2_SECONDARY_VALUE
The temperature of the O
2
cell.
O2_SECONDARY_VALUE_
The High and Low range limit values, the engineering units code, and the
RANGE
number of digits to the right of the decimal point for the O
2
cell
temperature.
O2_SENSOR_CAL_DATE
See FF-903 section 3.3.
O2_SENSOR_CAL_LOC
See FF-903 section 3.3.
O2_SENSOR_CAL_
See FF-903 sections 3.3 and 4.5.
METHOD
O2_SENSOR_CAL_WHO
See FF-903 section 3.3.
O2_SENSOR_INPUT
mV
The raw value of the O
2
sensor input.
O2_SENSOR_TYPE
See FF-903 section 3.3 and 4.3.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...