Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
5-4
Autocalibration using the Field Communicator - Fieldbus
To initiate an automatic calibration using 375 Field Communicator, perform
the following steps. Refer to Section 6, Field Communicator, for the
FOUNDATION Fieldbus menu tree.
1. From
Transducer
screen select
Methods
.
2. Select
OCX Calibration
from the
Methods
menu.
3. From the
OCX Calibration
menu, select the type of calibration desired:
Calibrate O2 Sensor
,
Calibrate Combustibles Sensor
, or
Calibrate Both Sensors
.
MANUAL CALIBRATION
If a unit is not equipped with calibration solenoids, a calibration must be
performed by an operator following prompts from the unit. Refer to the
following paragraphs for manual calibration.
Manual Calibration using the optional LOI
Use the following procedure to perform a manual calibration with the LOI. If
necessary, refer to the menu tree in Section 4, Using the LOI. Once the
manual calibration procedure is initiated at the LOI, a series of prompts will
appear giving instructions to the operator.
1. Use the right-pointing key to select the
CALIBRATION
first column
submenu.
2. From the
CALIBRATION
submenu use the right-pointing key to select
the
Cal Control
second column submenu.
3. From the
Cal Control
submenu use the right-pointing key to select the
third column
Start Cal O2
option.
4. Remain at the
Start Cal O2
option or use the down-pointing key to
select the
Start Cal COe
option or
Start Cal Both
option. (The following
sequence applies when
Start Cal Both
is selected.)
5. Use the right-pointing key to start the calibration. Turn on the low O
2
test
gas, when prompted by the
Flow Low Gas
message.
6. Press the right-pointing key when the low O
2
test gas is applied. The
calibration data changes as the calibration proceeds.
7. Press the right-pointing key when the low O
2
reading is stable. Turn off
the low O
2
test gas and turn on the high O
2
test gas as prompted by the
Flow High Gas
message.
8. Press the right-pointing key when the high O
2
test gas is applied. The
calibration data changes as the calibration proceeds.
9. Press the right-pointing key when the high O
2
reading is stable. Turn off
the high O
2
test gas. Press the right-pointing key to start the high O
2
gas
purge.
10. When the purge period expires, the LOI display reverts to the normal
operation display. If the calibration failed, the display will indicate an
alarm condition.
11. Press the right-pointing key to start combustibles calibration. Turn on
the CO test gas when prompted.
12. Press the right-pointing key when the CO test gas is applied. The
calibration data changes as the calibration proceeds.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...