Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
5-2
4. Scroll down to the last item
Use Solenoids
. If the unit is equipped with
calibration solenoids and timed automatic calibration is desired, select
Yes
.
5. Use the up-pointing key to select the item
O2 Out Tracks
. Select
Yes
or
No
to determine if updates to the O
2
lock value will take place.
6. Use the down-pointing key to select the item
COe Out Tracks
. Select
Yes
or
No
to define if updates to the COe lock value will take place.
7. Use the down-pointing key to select the item
O2 Cal Interval
. Enter the
amount of time in days and hours that is desired between automatic
calibrations.
8. Use the down-pointing key to select the next item
O2 Next Cal
. Enter
the amount of time in hours until the next automatic calibration. Select
the left-pointing key three times to move back to the LOI operating
display.
Autocalibration Setup using the Field Communicator - HART
NOTE
Automatic calibration is only available on units equipped with calibration
solenoids.
Use the following procedure to specify a time interval (in hours) at which the
OCX 8800 will automatically calibrate.
1. From the
DEVICE SETUP
screen, select
DETAILED SETUP
.
2. From the
DETAILED SETUP
screen, select
CAL SETUP
, then select
O2 CAL PARAMS
or
COE CAL PARAMS
.
3. If the unit is equipped with calibration solenoids and timed automatic
calibrations are desired, select
Solenoids
, then select
Y
es. Select
N
o to
disable the calibration solenoids.
4. Select
O2 Cal Intrvl
(O
2
calibration interval) and enter the desired time
in hours between automatic O
2
calibrations. Select
COE Cal Intrvl
and
enter the desired time between automatic COe calibrations. To disable
automatic calibration for O
2
and COe, enter 0 for both
Cal Intrvl
parameters.
5. If desired, the
O2 Next Cal Time
and the
COe Next Cal Time
(next
calibration time) parameters can be changed to synchronize a
calibration at a specific day or time.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
When setting automatic calibration times, CalIntrvl and NxtCalTm should be set so that O
2
and COe are NOT calibrated simultaneously.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...