Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
8-11
2
ALARM RELAY EVENTS
The OCX 8800 contains an alarm relay that can be configured to activate on
one of twelve different groups of events. These event groups, and the
conditions that trigger them, are listed in Table 8-2, Alarm Relay Event
Groups.
Table 8-2. Alarm Relay
Event Groups
In Calibration
Calibration in progress
Oxygen sensor heater temperature low
(O2 Temp Low)
O
2
Temperature Error
Heater Failure
Oxygen sensor heater temperature high
(O2 Temp Hi,
O2 Temp Very Hi)
RTD excitation current error
(Ref Curr Err)
Oxygen sensor heater could not reach final temperature
(O2 HTR Failure)
Combustibles sensor heater could not reach final temperature
(COe Htr Failure)
Sample Block sensor heater could not reach final temperature
(SB Htr Failure)
O Sensor Error
Oxygen sensor resistance high
(O2 Sensor R High)
Oxygen sensor disconnected
(O2 Sensor Open)
Calibration Failure
Calibration failed
(Cal Failed)
Calibration Warning
Calibration warning
(Cal Warning)
Board Temperature High
Electronics temperature maximum exceeded
(Board Temp Hi)
Any non-recoverable or heater relay off alarm
(O2 Temp Hi,
O2 Temp Very Hi, COe Temp Hi, COE Temp Very Hi,
SB Temp Hi, SB Temp Very Hi, O2 Htr Ramp Rate,
COe Htr Rmp Rate, SB Htr Ramp Rate, O2 TC Shorted,
Unit Failure
Sample Block
Temperature Error
COe Sensor
Temperature Error
Power Input Error
O2 TC Reversed, COe TC Shorted, COe TC Reversed,
SB TC Shorted, SB TC Reversed, ADC Failure,
ADC Ref Error, Board Temp Hi, EEPRM Chksm Fail,
Line Freq Error, Line Voltage Low, Line Voltage Hi,
Htr Relay Failed, Out Brd Fail)
Sample block heater temperature low
(SB Temp Low)
Sample block heater temperature high
(SB Temp Hi,
SB Temp Very Hi)
RTD excitation current error
(Ref Curr Err)
Combustibles sensor heater temperature low
(COe Temp Low)
Combustibles sensor heater temperature high
(COe Temp Hi,
COe Temp Very Hi)
RTD excitation current error
(Ref Curr Err)
AC power line frequency out of usable range
(Line Freq Err)
AC power line voltage below minimum
(Line Voltage Low)
AC power line voltage above maximum
(Line Voltage Hi)
All Alarms
Any alarm
Alarm Relay Event
Alarms/Conditions
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...