Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
3-4
Figure 3-2. OCX 8800 Defaults -
Fieldbus Electronics
INITIAL POWER UP
Allow adequate time (approximately 60 minutes) for the heaters to begin
operation and for the OCX 8800 to reach normal operating temperature on
power up. Normal operating temperature for the O
2
cell is 736°C. Normal
operating temperature for the combustibles cell is 300°C. The normal sample
line temperature is 170°C. During this time the eductor air solenoid will remain
closed so no sample is pulled through the analyzer. When the OCX reaches
operating temperature the solenoid will energize, eductor air will begin to flow,
and the unit will begin normal operation.
SET TEST GAS VALUES
Use Field Communicator or the optional LOI to set test gas values for
calibration. Refer to Section 4, Using the LOI or Section 6, Field
Communicator for more information.
Setting Test Gas Values with the Field Communicator
1. Use the 375 Field Communicator software to access the HART menu.
2. From the
DETAILED SETUP
menu, select
CAL SETUP
.
3. From the
CAL SETUP
menu, select
O2 CAL PARAMS
or
COe CAL
PARAMS
.
4. From
O2 CAL PARAMS
, select
O2 HIGH GAS
. Enter the percent O
2
used for the high O
2
test gas.
5. From
O2 CAL PARAMS
, select
O2 LOW GAS
. Enter the percent O
2
used for the low O
2
test gas.
6. From
COe CAL PARAMS
, select
COe Test Gas
. Enter the CO
concentration (ppm) used for COe test gas.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...