Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
9-36
Figure 9-28. Removal
of Tube Fittings
5
3
4
1.
Sensor Housing
2.
Electronics Housing
1
3.
Tube Fitting
4.
Tube Fitting (Type R)
5.
Tube Fitting (Type E)
3
2
3
3
Install Tube Fittings
1. Verify that the replacement tube fitting is identical to the item removed.
Special fittings are etched with code markings.
2. Apply pipe thread sealant (Loctite #567) to the mating threads of the
tube fitting. Do not apply sealant to the first turn of the external pipe
threads.
3. Install and tighten the tube fitting in the mating port of sensor or
electronics housing.
3
7
3
9
0
0
70