Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
3-9
CALIBRATION
TOLERANCE FEATURE
The calibration tolerance feature provides a mechanism to fail a calibration if
the calibration measurement does not fall within a specific tolerance of the
test gas value. The tolerance is preset within the OCX 8800 software and is
not user adjustable. The tolerance is different between the oxygen and
combustibles test gases. For oxygen, the calibration will fail if the measured
value differs by more than ±10% of the configured value. For combustibles,
the calibration will fail if the measured value differs by more than ±30% of the
configured value.
An OCX 8800 shipped from the factory must be configured before the
calibration tolerance feature can be implemented. This same process must
be performed any time a replacement card stack is installed.
Configuring the Calibration Tolerance Feature with the Field
Communicator - HART
1. Use the 375 Field Communicator or AMS software to access the HART
menu.
2. From the
DETAILED SETUP
menu, select
CAL SETUP
.
3. From the
CAL SETUP
menu, select
O2 CAL PARAMS
.
4. To enable the calibration tolerance feature for the oxygen calibration,
from the
O2 CAL PARAMS
, select
O2 Tol Check
. Select
On
to enable
the calibration tolerance feature.
5. Back out to the
CAL SETUP
menu, select
COe CAL PARAMS
.
6. To enable the calibration tolerance feature for the combustibles
calibration, from the
COe CAL PARAMS
, select
COe Tol Check
.
Select
On
to enable the calibration tolerance feature.
Configuring the Calibration Tolerance Feature with the Field
Communicator - Fieldbus
1. Use the 375 Field Communicator or AMS software to access the
Fieldbus menu.
2. From the
TRANSDUCER
block menu, select
O2 CAL/COe CAL
.
3. From the
O2 CAL/COe CAL
menu, select
O2 CAL SETUP/COe CAL
SETUP
.
4. From the
O2 CAL SETUP/COe CAL SETUP
menu, select
O2
Tolerance Check/COe Tolerance Check
. Select
Yes
to enable the
calibration tolerance feature.
Configuring the Calibration Tolerance Feature with the LOI
1. Use the "Z" pattern to enter the LOI menu tree.
2. From the
SYSTEM
menu, select
Calib Setup
.
3. From the
Calib Setup
menu, select the following:
O2 Tol Check
- Select
Yes
to enable the calibration tolerance feature
for the oxygen calibration.
Comb Tol Check
- Select
Yes
to enable the calibration tolerance
feature for the combustibles calibration.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...