Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
5-9
8. From the
SELECT ACTION
screen, select
START CAL/STEP CAL
to
continue calibration, select
ABORT CAL
to abort calibration or
EXIT
CAL
to exit calibration method. Select one from the list and press
ENTER
.
OCX: TAG NAME
SELECT ACTION
1. START CAL/STEP CAL
2. ABORT CAL
3. EXIT CAL
9. When the Calibration Status is at the
AppO2Low
step, switch on O
2
Low Gas. Verify the O
2
concentration measured matches the
O2 LOW
GAS
parameter in
Setup CAL
. Press
OK
when ready.
10. Select
START CAL/STEP
to start applying the O
2
Low Gas. The time to
apply the test gas is specified by the
Gas Time
.
11. The Calibration Status should automatically change to
FIowO2Low
and
then
ReadO2Low
for a period of time. During this period, if an attempt
is made to go to the next calibration step by pressing
OK
and selecting
START CAL/STEP CAL
, you will be prompted with
Operator step
command is not accepted at this time
.
12.
When ready, Calibration Status will stop at
AppO2Hi
. Switch off the O
2
Low Gas and switch on the O
2
High Gas. Verify the O
2
concentration
measured matches the
O2 HIGH GAS
parameter in
Setup
. Press
OK
when ready.
13. Select
START CAL/STEP CAL
to apply the O
2
High Gas. The time to
apply the test gas is specified by the
Gas Time
.
14. The Calibration Status should automatically change to
FlowO2Hi
, then
ReadO2Hi
, and then
ReadCOeLo
for a period of time. During this
period, if an attempt is made to go the next calibration step by pressing
OK
and selecting
START CAL/STEP CAL
, you will be prompted with
Operator step command is not accepted at this time
.
15. When ready, Calibration Status will stop at
AppCOeHi
. Switch off the O
2
High Gas and switch on the COe Gas. Verify the COe concentration
measured matches the
COe TEST GAS
parameter in the
Setup
. Press
OK
when ready.
16. Select
START CAL/STEP CAL
to start applying the COe Gas. The time
to apply the test gas is specified by the
Gas Time
.
17. The Calibration Status should automatically change to
FlowCOeHi
and
then
ReadCOeHi
for a period of time. During this period, if an attempt is
made to go the next calibration step by pressing
OK
and selecting
START CAL/STEP CAL
, you will be prompted with
Operator step
command is not accepted at this time
. The
Next Cal Step
command
is not accepted at this time.
When ready, Calibration Status will stop at
STOP GAS
. Switch off the COe
gas. Press
OK
when ready. Select
START CAL/STEP CAL
to start purging
gas. The time to purge gas is specified by
Purge Time
.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...