Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
7-58
Table 7-33. Input Selector Function Block System Parameters
Index
Number
Units
Description
ALERT_KEY
4
None
The identification number of the plant unit. This information may be used in the host for
sorting alarms, etc.
BLOCK_ALM
24
None
The block alarm is used for all configuration, hardw are, connection failure, or system
problems in the block. The cause of the alert is entered in the subcode field. The first
alert to become active will set the Active status in the Status parameter. As soon as the
Unreported status is cleared by the alert reporting task, another block alert may be
reported without clearing the Active status, if the subcode has changed.
BLOCK_ERR
6
None
This parameter reflects the error status associated w ith the hardw are or software
components associated with a block. It is a bit string, so that multiple errors may be
shown.
DISABLE_1
15
None
A Connection from another block that disables the associated input fromthe selection.
DISABLE_2
16
None
A Connection from another block that disables the associated input fromthe selection.
DISABLE_3
17
None
A Connection from another block that disables the associated input fromthe selection.
DISABLE_4
18
None
A Connection from another block that disables the associated input fromthe selection.
GRANT_DENY
9
None
Options for controlling access of host computers and local control panels to operating,
tuning, and alarm parameters of the block. Not used by device.
IN_1
11
Determined by
source
The connection input from another block. One of the inputs to be selected from.
IN_2
12
Determined by
source
The connection input from another block. One of the inputs to be selected from.
IN_3
13
Determined by
source
The connection input from another block. One of the inputs to be selected from.
IN_4
14
Determined by
source
The connection input from another block. One of the inputs to be selected from.
MIN_GOOD
20
None
The minimum number of good inputs The actual, target, permitted, and normal modes
of the block. Target: The mode to “go to” Actual: The mode the “block is currently in”
Permitted: Allow ed modes that target may take on Normal: Most common mode for
target.
OP_SELECT
22
None
Overrides the algorithm to select 1 of the 4 inputs regardless of the selection type.
OUT
7
EU of IN
The block output value and status.
OUT_UNITS
8
None
The engineering units of the output. Typically, all inputs have the same units and the
value is also the same.
SELECTED
21
None
The selected input number (1
–4).
SELECT_TYPE
19
None
Specifies selection method (see Block Execution).
STATUS_OPTS
10
None
Allows selection of options for status handling and processing. The supported status
option for the PID block is Target to Manual if Bad IN.
STRATEGY
3
None
The strategy field can be used to identify grouping of blocks. This data is not checked or
processed by the block.
ST_REV
1
None
The revision level of the static data associated w ith the function block. The revision
value w ill be incremented each time a static parameter value in the block is changed.
TAG_DESC
2
None
The user description of the intended application of the block.
UPDATE_EVT
23
None
This alert is generated by any change to the static data.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...