Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
8-6
O
2
cell degraded or failed:
Check O
2
cell impedance by reading the O2 Cell Imped value via the LOI (see
Figure 4-4, sheet 2), or the O2 Snsr R value via FOUNDATION fieldbus menu tree (see
O2 Slope Error (Slope <34.5 mV/Dec or >57.5 Figure 6-7, sheet 2). If cell impedance is zero, replace O
2
cell. If cell impedance is less
mV/Dec)
than 5000 ohms, check for cell housing ground fault. Repair ground fault. If cell
impedance is greater than 5000 ohms and no ground fault is indicated, replace O
2
cell.
O2 Constant Error (Constant not between -20
mV to +20 mV)
Flow calibration gas to the O
2
cell. Read the cell millivolt output. Plot the cell millivolt
output and the calibration gas O
2
concentration on the chart shown in Figure 8-2. If the
plotted values do not fall on the slope line shown in Figure 8-2, replace the O
2
cell.
Incorrect wiring between electronics and sensor
housings
Check all wiring between the electronics and sensor housings per Figure 8-3. Correct
wiring faults.
O
2
heater circuit wiring open
Check O
2
cell heater circuit for broken wire or loose connection per Figure 8-3 and
Figure 9-8. Repair broken wire or loose connection.
O
2
heater open
Check resistance of O
2
heater per Figure 9-8. Normal O
2
heater resistance is 62.5
Ohms. Replace O
2
heater if heater is open or has a large resistance.
Heater electronics failure
Check heater fuse F3 in electronics housing per Figure 8-1. If open, locate and correct
cause of overload. If F3 is not open, or if cause of overload cannot be found, replace
electronics package.
COe heater circuit wiring open
Check COe heater circuit for broken wire or loose connection per Figure 8-3 and
Figure 9-9. Repair broken wire or loose connection.
COe heater open
Check resistance of COe heater per Figure 9-9. Normal COe heater resistance is 97.7
Ohms. Replace COe heater if heater is open or has a large resistance.
Heater electronics failure
Check heater fuse F3 in electronics housing per Figure 8-1. If open, locate and correct
cause of overload. If F3 is not open, or if cause of overload cannot be found, replace
electronics package.
Sample block heater circuit wiring open
Check sample block heater circuit for broken wire or loose connection per Figure 8-3
and Figure 9-8. Repair broken wire or loose connection.
Sample block heater open
Check resistance of sample block heater per Figure 9-8. Normal sample block heater
resistance is 36.4 Ohms each (18.2 Ohms with both heaters in parallel). Replace
sample block heater if heater is open or has a large resistance.
Heater electronics failure
Check heater fuse F4 in electronics housing per Figure 8-1. If open, locate and correct
cause of overload. If F4 is not open, or if cause of overload cannot be found, replace
electronics package.
Calibration gas supply low or gas connection
leaking
Check calibration gas supplies and connections. Adjust gas pressure and flow.
Replenish low calibration gas supplies and tighten or repair loose or leaking
connections. When calibration gas supplies are adequate, recalibrate.
Table continued on next page
Sensor housing exposed to high wind and/or
extreme cold temperatures
If above probable causes are not causing the SB heater failure, install flange insulator
(PN 6P00162H01).
Cal Warning (Calibration warning)
Cal Failed (Calibration failed)
Probable Cause
Recommended Corrective Action
Electronics package failure
Replace electronics package.
O2 Htr Failure (Oxygen sensor heater could not reach final temperature)
Probable Cause
Recommended Corrective Action
ADC Failure (Voltage to digital conversion could not complete)
ADC Ref Error (Voltage to digital conversion not accurate)
Probable Cause
Recommended Corrective Action
COe Htr Failure (Combustibles sensor heater could not reach final temperature)
Probable Cause
Recommended Corrective Action
SB Htr Failure (Sample block heater could not reach final temperature)
Probable Cause
Recommended Corrective Action
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......