Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
3-5
Setting Test Gas Values with Fieldbus Communicator
1. Use the 375 Field Communicator software to access the Fieldbus menu.
2.
From the
TRANSDUCER
menu, select
O2 CAL.
3.
From
O2 CAL
menu select
O2 CAL SETUP.
4. From
O2 CAL SETUP
, select
O2 HIGH GAS
. Enter the percent O
2
used
for the high test gas.
5. From
O2 CAL SETUP
, select
O2 LOW GAS
. Enter the percent O
2
used
for the low test gas
6. From the
TRANSDUCER
menu select
COe CAL SETUP
.
7. From
COe CAL SETUP
, select
COe Test Gas
. Enter the CO
concentration (ppm) used for the COe Test Gas.
Setting Test Gas Values with the LOI
1. Use the "Z" pattern to enter the LOI menu tree.
2. From the
SYSTEM
menu, select
Calib Setup
.
3. From
Calib Setup
, select
O2 High Gas %
. Enter the percent O
2
used
for the high O
2
test gas.
4. Select the down arrow and the next selection will be
O2 Low Gas %
.
Enter the percent O
2
used for the low O
2
test gas.
5. Select the down arrow several times to display
COe Test Gas
. Enter the
CO concentration (ppm) used for COe test gas.
CALIBRATION
SOLENOIDS
The OCX 8800 can be provided with optional calibration solenoids for the
purpose of performing autocalibration. The solenoids are controlled by the
OCX 8800 software and automatically switch in the proper calibration gas
during the calibration cycle.
An OCX 8800 shipped from the factory with calibration solenoids must be
configured before autocalibration can be implemented. This same process
must be performed any time a replacement card stack is installed.
Configuring the Calibration Solenoids
with the Field Communicator - HART
1. Use the 375 Field Communicator to access the HART menu.
2. From the
DETAILED SETUP
menu, select
CAL SETUP
.
3. From the
CAL SETUP
menu, select
O2 CAL PARAMS/COe CAL
PARAMS
.
4. From the
O2 CAL PARAMS/COe CAL PARAMS
, select
Solenoids
.
Select
Yes
to enable the solenoids.
Configuring the Calibration Solenoids with the Field Communicator -
Fieldbus
1. Use the 375 Field Communicator to access the Fieldbus menu.
2. From the
TRANSDUCER
block menu, select
O2 CAL/COE CAL
.
3. From the
O2 CAL/COE CAL
menu, select
O2 CAL SETUP/COE CAL
SETUP
.
4. From the
O2 CAL SETUP/COE CAL SETUP
, select
Solenoids
. Select
Present
to enable the solenoids.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...