Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
9-5
OCX with
Remote Electronics
Remove Sensor Housing
1. Turn off power to the system.
2. Shut off the test gasses at the cylinders and close the instrument air
valve.
3. Disconnect the calibration gas, reference air, eductor air, and dilution air
lines from the sensor housing, Figure 9-3.
Figure 9-3. OCX with
Remote Electronics
Duct
Stack
Sensor
Housing
Heater
Power Cable
Signal Cable
Signal
Outputs
AC Power
Input
Electronics
Housing
Instrument Air
(Reference Gas)
High O
2
Test Gas
Low O
2
Test Gas
CO Test Gas
3
7
3
9
0
0
44