Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
7-11
1. Run "Transmitter Options" method.
2. In the "Transmitter Options" menu, select either "Simulate Switch" or
"PWA Simulate".
3. Select "Enable" to enable the simulation feature or "Disable" to disable
the simulation feature.
4. Once the method is complete, select the "Simulate PWA" tab in the
Resource Block, Figure 7-4. If the simulation is enabled, the "PlantWeb
Alarm Simulate" parameter is configurable otherwise it is read-only.
5. Now select Simulation on/off from the "Simulate PWA" screen. When
Fieldbus Simulation is on, the "Simulation Switch" LED is illuminated.
When PWA simulation is on, the "PWA Simulate" LED is illuminated.
Figure 7-4. Simulate PWA Screen
6. If PWA Simulation is on, all PWA active parameters and Resource and
Transducer Block status parameters are configurable. Otherwise they
are read-only.
7. To simulate PlantWeb alerts, use the "Simulate PWA" screens in the
Resource Block. To simulate block alarms, use "Simulate TB
Temperature Sensor - Temperature Status", "Simulate TB Temperature
Status", or "Simulate TB Calibration Device - FF Status" in the
Transducer Block.
8. Select "Device Diagnostics" to view the active PWA alarms.
9. When "Device Diagnostics" is selected, press the "Status" button to see
the Resource or Transducer Block detailed status displays.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...