Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
9-31
Remove Solenoid Valves
1. Disconnect solenoid leads from mating terminal connector.
2. Remove top nut of solenoid valve (15 or 16, Figure 9-24).
3. Remove the solenoid coil assembly and washer.
4. Unthread and remove solenoid valve base.
Remove EMI Filter and Terminal Block
1. Disconnect EMI filter wiring (Figure 9-25) at terminal block (3).
2. Disconnect EMI filter wiring at AC power input terminal block on
electronic stack.
3. Unbolt and remove EMI filter (1) from electronic stack.
4. Remove ground wire (2) from terminal block (3).
5. Unbolt and remove terminal block (3) from electronic stack.
Figure 9-25. Removal/
Installation of EMI Filter
Electronics Housing
Assembly
Install EMI Filter and Terminal Block
1. Install replacement EMI filter (1, Figure 9-25) and/or terminal block (3)
on electronic stack.
2. Refer to wiring details in Figure 9-25. Connect EMI filter wiring and
ground wire (2) at terminal block (3).
3. Connect EMI filter wiring at AC power input terminal block on electronic
stack.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...