Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
7-26
Convert
Cutoff
Filter
Alarm
Detection
PV
Status
Calc.
OUT (mode in man)
OUT (mode in auto)
PV
63% of Change
FIELD_VAL
Time (seconds)
Figure 7-6. Analog Input
Function Block Schematic
Analog
Measurement
HI_HI_LIM
HI_LIM
LO_LO_LIM
LO_LIM
EL
ALARM_HYS
LOW_CUT
ALARM_TYPE
OUT_D
OUT
SIMULATE
L_TYPE
PV_FTIME
MODE
FIELD_VAL
OUT_SCALE
XD_SCALE
IO_OPTS
STATUS_OPTS
NOTES:
OUT = block output value and status.
OUT_D = discrete output that signals a selected alarm condition.
Figure 7-7. Analog Input
Function Block Timing Diagram
PV_FTIME
Filtering
The filtering feature changes the response time of the device to smooth
variations in output readings caused by rapid changes in input. You can adjust
the filter time constant (in seconds) using the PV_FTIME parameter. Set the
filter time constant to zero to disable the filter feature.
Signal Conversion
You can set the signal conversion type with the Linearization Type (L_TYPE)
parameter. You can view the converted signal (in percent of XD_SCALE)
through the FIELD_VAL parameter.
3
8
7
4
0
0
8
1
3
8
7
4
0
0
8
0
Access
Analog
Meas.
CHANN
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...