Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
7-16
Parameter Mnemonic
Valid Range
Units
Description
O2_SETPOINT
°C
O
2
heater temperature set point.
O2_SLOPE
34.5-57.5
mV/Decade
The O
2
calibration slope.
O2_T90
0-300
Seconds
The amount of time that the O2 process variable will take to reach 90% of
the actual process variable.
O2_TEMP_MAX
°C
The highest O
2
temperature read since power on.
O2_THERMOCOUPLE_
mV
The raw value of the O
2
temperature input.
INPUT
O2_TIME_TO_NEXT_CAL
0-9999
Hours
The time until the next automatic calibration of the O
2
sensor.
O2_TOL_CHECK
0: No
– 1: Yes
O
2
calibration gas tolerance check.
OPERATING_MODE
Enumerated
Device Operating Mode. See Table 7-8.
PCD_COUNTER
Power cycle drop counter.
PCDC_ENABLE
0: No
– 1: Yes
Enable/disable power cycle drop detect.
PCN_COUNTER
Power cycle noise counter.
SB_HTR_DUTYCYCLE
Sample Block heater duty cycle.
SB_SETPOINT
°C
Sample Block heater temperature set point.
SB_TEMP
°C
The temperature of the sample line.
SB_TEMP_MAX
°C
The highest Sample Block temperature read since power on.
SB_THERMOCOUPLE_
mV
The raw value of the sample line temperature input.
INPUT
SENSOR_HOUSING_TEMP
°C
Sensor Housing temperature (Valid only with Type 3 sensor).
SENSOR_HOUSING_
mV
Sensor Housing CJC voltage (Valid only with Type 3 sensor).
TEMP_INPUT
SENSOR_HOUSING_
°C
The highest Sensor Housing temperature read since power on.
TEMP_MAX
SENSOR_HOUSING_TYPE
0, 1, 2, 3
Enumerated
This is the Sensor Housing type setting through the DIP switch.
(0=Type 1, 1=Type 2, 2=Type3, 3=Invalid)
SOLENOIDS_PRESENT
0: No
– 1: Yes
Enumerated
This determines whether a calibration cycle will automatically step
through, turning solenoids on and off to switch test gas, or wait for an
operator to manually switch gases and acknowledge.
ST_REV
See FF-891 section 5.3.
STATS_ATTEMPTS
Total number of messages sent to the transducer a/d board.
STATS_FAILURES
Total number of failed a/d board message attempts.
STATS_TIMEOUTS
Total number of timed out a/d board message attempts.
STRATEGY
See FF-891 section 5.3.
TAG_DESC
See FF-891 section 5.3.
TRANSDUCER_
See FF-903 section 3.3.
DIRECTORY
TRANSDUCER_TYPE
See FF-903 sections 3.3.
UPDATE_EVT
See FF-891 section 5.3.
VERIFY_STATE
0 through 5
Enumerated
The current state of the Calibration Check. (0=Idle 1, 1=Flow High O
2
,
2=Flow Lo O
2
, 3=Flow High COe, 4=Purge Gas, 5=Done)
VERIFY_STATE_STEP
0, 1, 2,.3, 6
Enumerated
Initiates a calibration verify of O
2
or Combustibles gas. (0=Start Flow High
O
2
, 1=Start Flow Lo O
2
, 3=Start Flow High COe, 3=Purge Gas, 6=No
Effect)
VERIFY_STATE_TIME
Seconds
Time remain for the current Calibration Check state.
XD_ERROR
See FF-903 section 3.3.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...