Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
1-13
Blowback Air (optional)
Clean, dry, instrument-quality air (20.95% O
2
), regulated to
≥ 60 psi
(413 kPa) or greater and ambient temperature of
≥ 0 °F (-18 °C)
Specifications
Sensors Housing
Type 4X, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports (when
reference air vents are routed to a dry area).
Electronics Housing
Type 4X, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports (when
reference air vents are routed to a dry area).
Certifications
Electrical Noise
EN 61326-1, Class A
Line Voltage
Universal 100 to 240 VAC ±10%, 50 to 60 Hz, no switches or
jumpers required, 3/4-14 NPT conduit port
Pollution Degree
2
Over Voltage Category
II
Relative Humidity
5 to 95% (non-condensing)
Isolated Output
Oxygen
4-20 mAdc, 950 ohm maximum, with HART or FOUNDATION
fieldbus capability only
Combustibles
4-20 mAdc, 950 ohm maximum (Not present with FOUNDATION
fieldbus)
Alarm
Alarm output relay - dry contact, form C, 30mA, 30VDC capacity
Power Consumption
750 W maximum
NOTE
All static performance characteristics are with operating variables constant. Specifications subject to change
without notice.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...