Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
7-10
Alerts
Alarms
AMS
Tab
What does the
alert indicate?
Recommended
Action
Description
Line Input Out of Transducer
Device
This alert
Check line input
The device power supply continuously monitors
Range
Block:
Alerts
indicates that the power for proper
the line input. Measured variations in the line
line input power
voltage and
input power are used to compensate the sensor
1) Line
to the device is
frequency.
heater control and check for faulty line conditions.
Frequency
outside the
Refer to Section 8: Troubleshooting for details.
Error
proper operating
2) Line Voltage
limits.
Low
3) Line Voltage
High
Inter Board
Transducer
FF/
This alert
Verify device is
There is no communication possible between the
Communication
Block:
Device
indicates a
powered. Check the Fieldbus Output Board and
the device’s computer
Failure
Alerts
communication
installation of the
board.
Inter Board
failure between
Fieldbus Output
Communication
the FF board and Board on its carrier
Failure
the device.
board. If the above
are OK, replace the
carrier board and/or
Fieldbus Output
Board.
Simulate Active
FF/
This alert occurs
The PWA active parameters can now be written.
Device
when the PWA
The resource block detailed status parameters
Alerts
simulate mode is
and the internal alerts in the Transducer Block
active.
where the PWA active alarms originate can also
be written.
PWA Simulate
Setting PWA_SIMULATE to ON allow simulating the PWA parameters,
FAILED_ACTIVE, MAINT_ACTIVE and ADVISE_ACTIVE. “Allowing
Sim
ulating” means that these parameters get write permission and the host’s
written value is the only one which is used for parameter’s read back value.
The data which comes from the OCX 8800 is not used in this case.
Fieldbus/PWA Simulate
Fieldbus simulation and PWA simulation can be enabled and disabled using
the DD method. Use the "Transmitter Options" method to enable/disable the
Fieldbus or PlantWeb Alerts simulation.
Selecting “Fieldbus simulation” enables both the Fieldbus function block
simulation and
PWA simulation. Selecting “PWA simulation” enables
PlantWeb Alert simulation only.
The simulation feature should not be used for normal operations. This feature
is to be used by authorized personnel for testing or demonstration purposes
only.
Configure Simulation from AMS
Use the following procedure to configure PWA simulation using AMS.
Do not use the PWA Simulate feature for normal operations. When used improperly the
Simulate feature can alter, disable, or activate device alarms.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...