Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
3-7
Blowback Purge Time
- Length of time after blowback is complete
before oxygem/combustibles readings are considered valid (Set as
required by the application).
Initiate Blowback
- Initiates a blow back event manually.
Configuring Blowback with the LOI
1. Use the "Z" pattern to enter the LOI menu tree.
2. From the
SYSTEM
menu, select
Blow Back
.
3. From the
Blow Back
menu, select
Blow Bk Enable
. Select
Yes
to
enable blowback. Also set the following parameters:
Blow Bk Intrvl
- Length of time between blowback events. Range is 0
to 32,000 minutes. Default is 60 minutes. 60 minutes is recommended.
Blow Bk Period
- Length of time blowback in activated. Range is 1 to 5
seconds. Default is 2 seconds. 5 seconds is recommended.
Blow Bk Purge
- Length of time after blowback is complete before
oxygem/combustibles readings are considered valid. Range is 0 to 500
seconds. Default is 88 seconds. Set as required by the application.
Force Blow Bk
- Initiates a blow back event manually.
CALIBRATION VERIFY
FEATURE
The calibration verify feature flows one or more calibration gases to verify the
analyzer is reading correctly. The calibration verify feature flows each
calibration gas on demand to verify calibration, but does not change the slope
or constant of the current calibration. This function uses the same gas flow
and purge times from the basic calibration setup.
The calibration verify feature is only valid if the OCX 8800 is supplied with
calibration solenoids and the solenoids have been activated.
Performing a Calibration Verify with the Field Communicator - HART
1. Use the 375 Field Communicator or AMS software to access the HART
menu.
2. From the
DEVICE SETUP
menu, select
DIAG/SERVICE
.
3. From the
DIAG/SERVICE
menu, select
CALIBRATION
.
4. From the
CALIBRATION
, select
CAL VERIFY
. Select
Verify
Calibration
. From this menu, select the functions as follows:
Flow High O2 Gas
- Flows the high O2 test gas for the time specified in
the calibration setup.
Flow Low O2 Gas
- Flows the low O2 test gas for the time specified in
the calibration setup.
Flow High COe Gas
- Flows the COe test gas for the time specified in
the calibration setup.
During the Calibration Verify function the analog output signals will track the oxygen and
combustibles readings. To avoid a potentially dangerous operating condition, the OCX
8800 must be removed from the automatic combustion control loop before performing the
Calibration Verify procedure.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...