Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
9-34
SOL 3
#1
#1
Figure 9-27. Solenoid
Power Terminals
CO Test Gas
TOP VIEW
High O
2
Test Gas
Low O
2
Test Gas
Instrument Air
Install Cover
1. If removed, install cover gasket (4, Figure 9-24). Screw cover (3) onto
electronics housing. Tighten cover firmly. Align locking clip (2) with gap
between cover ribs.
2. Loosen screw (1) and slide locking clip (2) fully into gap between cover
ribs. Retighten screw (1).