Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
9-2
OCX with
Integral Electronics
Remove OCX 8800
1. Turn off power to the system.
2. Shut off the test gases at the cylinders and shut off the instrument air.
3. Disconnect the test gas and instrument air lines from the electronics
housing, Figure 9-1.
Figure 9-1. OCX with
Integral Electronics
Stack
Adapter
Plate
Duct
Signal Outputs
(Twisted Pairs)
AC Power Input
Sensor
Housing
Electronics
Housing
Instrument Air
(Reference Gas)
High O
2
Test Gas
Low O
2
Test Gas
CO Test Gas
4. Remove the cover from the electronics housing to expose the
electronics housing terminal blocks, Figure 9-2.
5. Disconnect and remove the power leads from the AC power input
terminal block and remove the ground lead from the ground stud.
6. Disconnect and remove the O
2
and COe signal leads from the 4-20 mA
signal output terminal block.
7. If used, disconnect and remove the external relay leads from the alarm
output relay terminal block.
8. Disconnect and remove customer power and signal wire conduits and
wiring from the electronics housing.
9. Remove insulation to access the sensor housing mounting bolts. Unbolt
the OCX 8800 from the stack and take it to a clean work area.
10. Allow the unit to cool to a comfortable working temperature.
3
7
3
9
0
0
4
3
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...