COe sensor degraded
COe Slope Error (Slope not between 200
ppm/Ohm to 4,500 ppm/Ohm)
COe Constant Error (Constant not between
-100 Ohms to +100 Ohms)
Replace COe sensor or increase warning level (
SYSTEM
,
CALIB SETUP
,
COe Slope
Warn
via HART.)
Low sample gas flow in sensor housing due to
flow path plugging
Check the following portions of the flow path for plugging:
• blowback filter
• in-situ filter
• eductor outlet path
Calibration Gas concentration incorrect or gas
bottle empty
O
2
TOL Error (not within 10% of configured O2 test gas).
COe TOL Error (not within 30% of configured CO test gas).
Table continued on next page
Figure 8-2. Oxygen and Cell
Output
9
7
5
0.9
0.7
0.5
-20
0
20
40
60
80
100
120
ANALYZER OUTPUT (MILLIVOLT)
Cal Warning (Calibration warning)
Cal Failed (Calibration failed)
Probable Cause
Recommended Corrective Action
O
X
Y
G
E
N
CO
NCEN
T
R
A
T
IO
N
(%)
3
7
3
9
0
0
5
1
20
10
(23)
8
6
4
3
2
1
(73)
0.8
0.6
0.4
0.3
0.2
0.1
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...