6100-Gfast Multifunction Network Analyzer Operation Manual
74
© 2020 Radiodetection Ltd
Threshold Settings
button opens a new page
that allows you to set the
Stressed Balance
Threshold
for the test.
Result
displays the measurement value for
Stressed Balance
and pass/fail status
dependent on the
Good/Marginal Balance
threshold settings.
Test Status
displays the current state of
measurement or an error message.
To select the Stressed Balance parameter values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options.
4. Press
to confirm the value.
Threshold Settings
The
Threshold Settings
page allows you to set the
Good/Marginal Balance
threshold values for the
Stressed Balance
test. The threshold is a
programmable value you can use as an on-screen
indicator in each test when a measured value is too
high or too low.
Type
is the threshold limit and is fixed to
Maximum
.
Enabled
is a flag to use the current threshold:
either
Yes
or
No
.
Note: Disabling
Good Balance
will also disable the
Marginal Balance
threshold, even if you
Enabled
it.
To set threshold values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
Result Details
The
Result Details
read-only page displays the
Stressed Balance
results and the pass/fail status.
Good values are lower (
≤ 20) compared to the VF
(longitudinal) Balance test (
≥ 60).
Test Group
is the overall test group of the
selected saved results file, for example
Multimeter
.
Test Identification
is the test type of the
group, for example
Stressed Balance
.
Test Modes
identifies the manner in which the
test was run, for example
Continuous
.