6100-Gfast Multifunction Copper Tester Operation Manual
© 2020 Radiodetection Ltd
147
o
Export
button confirms if the
Report
File Name
is saved successfully. If a
USB device is not inserted, the
following warning dialogue box
appears:
USB device not present
.
To set parameter values and save results:
1. Press the left/right and up/down arrow keys to
highlight the desired parameter.
2. Press to display the list or select the value.
3. Press the up/down arrow key to highlight the
desired value.
OR
4. Press the left arrow key to erase the existing
value, and then use the alphanumeric keypad
to enter a value. To cancel the entry, press
.
5. Press to accept the value.
Multimeter 1 Result
The
Multimeter 1 Result
page displays the following
results and pass/fail status:
AC-DC Voltage
- volts and frequency (
Hz
)
values for each continuous AC/DC
Pair
.
AC-DC Current
-
DC
current,
AC
current and
frequency (
Hz
) values for each wire
Pair
.
Resistance
for each wire
Pair
and
Length
for
T/R/A/B
.
Multimeter 2 Result
The
Multimeter 2 Result
page displays the following
results and pass/fail status:
Resistive Balance - R/B-Leg, T/A-Leg, Loop
resistance, and pass/fail status and values in Ω
and %.
Isolation
- pass/fail status,
Resistance
and a
Soak Counter(s)
for each wire
Pair
. If your
unit has the HIVOLT (high voltage) software
option configured, it enables
Isolation
Resistance
testing up to 500 V. Default is
inherited from
Multimeter Isolation
test.
Capacitance/Opens
and
Length
for each wire
Pair
and overall
Capacitive Balance
.
Noise Result
The
Noise Result
page displays the following results
and pass/fail status: