6100-Gfast Multifunction Copper Tester Operation Manual
© 2020 Radiodetection Ltd
103
Current Group
is the current test group name,
for example
Frequency Tests
.
Current Test
displays the test being
measured. Here you can switch to another test
from the given list. However, this action will
terminate the ongoing measurement test.
Threshold Settings
button opens a new page
that allows you to set the
VF/AC Balance
threshold.
Result
displays the pass/fail status and
Active
VF/AC Balance
value.
Test Status
displays the current state of
measurement or an error message.
Note: The selected
Standard
in
Setup/Application
Settings
determines the name of the frequency
test -
ANSI: VF Balance
,
ITU: AC Balance
.
To select the parameter values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options.
4. Press
to confirm the value.
VF/AC Balance Active Threshold
Settings
The
Threshold Settings
page allows you to set the
Active VF/AC Balance Threshold
values for the test.
VF/AC Balance
sets the active balance
threshold.
Type
is the threshold limit: either
Maximum
or
Minimum
.
Enabled
is a flag to use the existing threshold:
either
Yes
or
No
. For example, setting the
threshold
Enabled
to
No
will disable pass/fail
threshold.
To set threshold values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
Active Result Details
The
Active Results Details
read-only page displays
the
VF/AC Balance
value and the pass/fail status.