6100-Gfast Multifunction Network Analyzer Operation Manual
146
© 2020 Radiodetection Ltd
OR
3. Press on a value to open a list box of
options.
4. Press to confirm the value.
5. Press to confirm and close your
selections.
Upload/Save Result
The
Upload/Save Result
tab allows you to
save/upload/export
Auto Test
results, using the upload
method (Wi-Fi, or Ethernet) previously selected in
System Settings\Upload Setup
.
The tab displays a screen where the following
information can be entered and functions performed:
Identification Fields
o
User Name/User ID
display the values
set up in the Application
Settings/Identification tab. You can edit
this field using the editor screen.
o
Contractor Name
allows you to enter
the name of a contractor used to
perform the tests.
o
Test From/To
display the values set
up in the
Application
Settings/Identification
tab. You can
edit this field using the choices from
the list box.
o
Job ID
allows you to create/edit a
unique identifier for the task.
o
Customer Name
is the name of the
customer for which the test was run.
o
Circuit ID
allows you to create/edit a
unique identifier for the circuit under
test.
o
Comments
allow you to add any
relevant information.
Save Result
o
Select Result File
button displays a
list of existing file names where you
can save your results.
o
Result File Storage Location
gives
you the option of storing results either
in the 6100-Gfast internal memory or
USB device. The default selection
comes from Application
Settings/General tab.
o
Result File Name
previews the result
file name which consists of the
Circuit
ID
, date and time as configured in
System Settings/Date and Time
, and
version number if there are multiple
versions with the same filename.
o
Upload & Save
button initializes the
result file upload and confirms if the
Upload File Name
is saved
successfully.
o
Save
button confirms if the
Result File
Name
is saved successfully.
Export Report
o
Report File Storage Location
is set to
USB automatically.
o
Report Format
is
HTML, MHTML
, or
XML
.
You can generate and export HTML,
MHTML, or XML reports from any
saved test result.
Reports include:
General Information
including
Pass/Fail
status.
Identification
Profile/Result Summary
Blank entries for your
Signature
and
the
Date
.
o
Report File Name
previews the result
file name which consists of the
Circuit
ID
, date and time as configured in
System Settings/Date and Time
, and
version number if there are multiple
versions with the same filename.