Troubleshooting
This chapter describes items to check if the system does not function as you expected, and meaning and handling of the
warning messages.
Incorrect use might prevent this system from providing primary performance even though the system is not damaged. If a
symptom such as those in the following table occurs, take action as described in the table before requesting repair.
If a problem that is not listed in the table occurs or if a problem that occurred is not resolved by taking action as described
in the table, disconnect the power cable of the device, and then contact your nearest Nikon representative.
Note that the tables in the following subsections cover only issues on operating the DS-L4 and a microscope camera. For
issues on using the DS-L4 connected to the Nikon microscope, refer to the separate volume of this manual, "Microscope
Operation".
16.1
Power Supply
Symptom
Possible Cause
Action
The power does not come on.
The AC adapter is not connected.
Connect the DC output cable of the cradle's AC adapter
to the DC IN connector of the cradle.
The power supply cord is connected
incorrectly to the AC adapter.
Use the specified power supply cord to plug the AC inlet
of the AC adapter into the AC outlet.
The power is off.
Press the power button to turn on the DS-L4 power.
The DS-L4 is hot when touched.
The DS-L4 is used in a hot or confined
space.
Use the DS-L4 within the operating environment
described in "Chapter 20
Main Specifications
".
If an unusual odor ( burning, etc.) occurs, turn off the
power, disconnect the power plug of the AC adapter,
and request repairs.
16.2
Image Output
Symptom
Possible Cause
Action
Images are not displayed on the
display panel.
The subject image is not incident in the
optical device.
Set the subject image to be captured correctly.
The optical path of the optical device is
not set in the camera side.
Set the optical path correctly.
The illumination or exposure setting of
the optical device is improper.
Use illumination appropriate for the subject to adjust
exposure properly.
"No Camera" is displayed.
The microscope camera is not
connected.
Turn off the DS-L4, and using a USB 3.0 cable,
securely connect the camera and the DS-L4.
Images are not displayed on the
external display.
Cables are connected incorrectly.
Connect a commercially available cable between the
cradle's analog display connector or DisplayPort
connector and the external display's video input
connector.
(The analog display connector and the DisplayPort
connector cannot be used concurrently.)
The external display does not support
the video format of the DS-L4.
Use an external display that supports the 1920 x 1200
video format.