Chapter 16
Troubleshooting
176
O
ther
Informat
ion
16.6
System Operation
Symptom
Possible Cause
Action
The mouse or keyboard does not
work properly or does not work
at all.
Operation of the component has not
been verified by Nikon.
Use a component whose operation has been verified by
Nikon.
(For USB) USB hubs are cascaded.
Use only a single-tier hub.
(For Bluetooth) Battery is low.
Charge or replace the battery.
(For Bluetooth) Pairing has not finished
yet.
Perform pairing.
The keyboard does not work
properly or does not work at all.
The DS-L4 language setting is not
compatible with the keyboard.
To use a Japanese keyboard, set the DS-L4 language
to Japanese. To use an English keyboard, set the
DS-L4 language to English.
16.7
Network
Symptom
Possible Cause
Action
The system cannot access the
network.
(If a wired connection is used) The
network cable is not connected correctly.
Connect it correctly.
The network cable is not correct.
Make sure that the category of the network cable is
correct for the network.
To connect the DS-L4 via a hub, use a straight cable.
To connect the DS-L4 directly to the network port of the
PC without using a hub, use a crossover cable.
The network settings are not correct.
Specify the network settings correctly.
Take particular note of the IP address or network
enablement settings. To perform data transmission and
reception between different network groups, be sure to
specify the gateway address.
DHCP is set in an environment where no
DHCP server exists.
For IP addresses to be assigned dynamically via the
DHCP protocol, a DHCP server is required in the
network.
(If a wireless connection is used)
Connection to a wireless LAN access
point is not established.
Use DS-L4 Administration Tool to specify wireless LAN
or SoftAP settings.