FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
57
Signal Test
The signal test has two functions:
•
Checking the state of the overall system: This function is
typically performed using the yellow test plate with the
respective plate format for your assay and can be accessed
through the
Yellow Plate Signal Test
command in the main
menu. Outlined in the table below are the default settings used
when performing the yellow plate signal test. These settings
cannot be saved to a protocol for use at a later time.
•
Checking initial fluorescence of a plate prior to running an
assay: This function is typically performed to evaluate the assay
plate prior to running an experiment and can be accessed
through the
Protocol Signal Test
command in the main menu.
Settings outlined in the table below will default to the settings in
the currently open protocol when performing the protocol signal
test. These settings can be saved to the open protocol for use at
a later time.
Set Remote/
Manual Mode
A toggle that switches the instrument between manual and
remote modes.
Set Remote Mode
—Enables you to integrate a third-party
robotics system with FLIPR
®
Tetra System. See
Robotic Integration on page 221
ScreenWorks automatically goes into
Remote Mode
if you
open
SynchroMax Automation
, however
Remote Mode
is not used to initiate communication with the SynchroMax
ET.
Set Manual Mode
—Disables SynchroMax ET or third-party
robotics control. All instrument commands must be initiated
through ScreenWorks.
SynchroMax
Automation
Opens the
SynchroMax
dialog for configuration of plate
handling with the SynchroMax ET. See
Reset
Reinitializes the system to clear any fatal errors as
designated in red at the bottom of the
Instrument Status
panel.
Clear Error
Clears minor system errors as designated in yellow at the
bottom of the
Instrument Status
panel.
Calibration
Opens the
Calibration
dialog where
Flat Field Calibration
Calibrations can be performed. See
Refresh
Configuration
Refreshes the instrument configuration. Use this command
after hardware settings have changed.
Item
Description
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...