ScreenWorks Software Overview
108
0112-0109 H
Transfer Fluid Process
This process enables you to program reagent or cell additions and to
execute them in conjunction with other processes.
The default protocol has one
Transfer Fluid
process.
A blue
Transfer Fluid
icon indicates that the process occurs in series
with other blue-colored processes in the protocol. A purple icon
indicates that the execution of the process is synchronized to a Read or
Mix Fluid
process (which will also have a purple icon).
The
Transfer Fluid
process has different options for pin tools and
standard pipettors. In each case you must configure aspiration and
dispense of compound separately.
For a standard pipettor you can configure more than one aspiration or
dispense (but not both) within the process, with the appropriate
‘multiple’
Fluid Transfer Type
.
With a standard pipettor selected, aspiration and dispense steps are
configured by selecting the appropriate row in the aspiration or
dispense table and pressing the
Edit
button to open a configuration
dialog. With the Pin Tool, aspiration and dispense configuration is done
directly in the
Transfer Fluid
page.
To enable additional dispense steps with a standard pipettor, select the
multiple dispense option in the
Fluid Transfer Type
. Then configure
the first dispense with a volume less than the volume aspirated. An
additional dispense is automatically enabled for the remaining volume.
Similarly, for multiple aspirates, configure the first aspirate with less
volume than the dispense.
Note:
The entire volume of fluid aspirated is always dispensed.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...