FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
201
103: “Invalid start of
command.”
Start of command is
not valid. See
appended string on
the error message for
details.
Contact Technical Support
at
+1-800-635-5577.
104: “Command invalid,
system busy.”
The system cannot
act upon the
command, as the
system is busy. See
appended string on
the error message for
details.
Contact Technical Support
at
+1-800-635-5577.
“Data buffer full”.
The data buffer is full.
See appended string
on the error message
for details.
Contact Technical Support
at
+1-800-635-5577.
106: “Option not installed,
command unavailable”.
Pipettor head, LED,
Emission Filter or
TETRAcycler
™
is not
installed.
Install option you want to
use.
Change protocol to use
available instrument
options.
107: “Command failed”.
General failure.
Contact Technical Support
at
+1-800-635-5577.
108: “Command failed,
system in an error state".
Minor communication
error occurred.
Select
Clear Error
command from
Instrument
drop-down
menu.
109: “Command failed,
system not initialized “.
Communication failed
between software and
instrument.
Select
Reset
from the
Instrument
drop-down
menu.
Turn the instrument
power switch OFF then ON
to cycle power.
“Password does not
match”.
User password is
incorrect.
Contact your instrument
administrator for the
appropriate user name
and password.
Table 9-3
General Troubleshooting (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...