FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
145
To remove the existing plate hold-down:
1.
Clear the system of plate or tips that may interfere with the
position from which you are removing the plate hold-down
device.
2.
Use a Phillips #2 or #3 screwdriver to loosen and remove the
screws from the center of the devices.
3.
Remove the plate hold-down device.
To install or reposition the plate hold-down:
1.
Place the device in its proper orientation so that the numbers
appear upright when viewed looking through the main user
access door.
2.
Insert one of the flat head screws that was originally securing
the device or one included in the accessory kit through the hole
in the device and into the screw threads on the left side of the
plate-retaining pockets. Tighten the screw firmly. The device is
self locating therefore no additional alignment will be necessary.
3.
If plates or tip blocks with different flange heights are to be used
in the same position, use the device specified for the largest
flange thickness.
4.
Test the plates or tip blocks to be used by inserting them into
the plate pockets at the location in which they will be used.
5.
Observe that when the plate is released it is pushed to the left
under the retaining device. It should come to rest against the
edge locating surfaces on the left and nearest the door.
6.
Pull up on the plate gently. A definite retaining force should be
felt. It should be noted that a sideways movement during lifting
can cause the plate to slide out from under the retaining device
and be lifted upward without retention. This does not happen
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...