Robotic Integration
270
0112-0109 H
Instrument errors
Instrument errors are errors that are generated by executing a
command. These errors are detected during the execution of the
command by the instrument. They will be reported in the status
response in the function line. These errors consist of a three-digit
numeric code.
Instrument errors fall into two general categories, recoverable and fatal
errors. When fatal errors occur the instrument will not be able to
continue to do any operation. It will abort any pending operations.
When recoverable error occurs with user interaction instrument will be
able to continue on.
The
status
command will return ERRORCODE or FATALCODE.
•
Error codes 100–199 ask you to call Tech support.
•
Error codes 200–299 are instrument recoverable errors.
•
Error codes 300–399 are instrument non-fatal errors.
•
Error codes 400–499 are instrument fatal errors.
•
Error codes 500–599 are instrument diagnostic errors.
A full list of these errors can be found in
Appendix A—Remote Interface Revision History
V1.0
Creation
V1.1
Added commands:
•
version
•
loadtips
•
unloadtips
•
cyclecameratemp
•
tempcontrolonoff
•
washtips
Error code response will include text description after code. for
example
,
“c45, folder does not exist”.
V1.2
Added Error Code:
C01–Instrument is not in
Remote Mode
. Only
status
command can
be sent while instrument is in
Manual Mode
.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...