SynchroMax™ Automation
The
SynchroMax Automation
command is only enabled if you have a
SynchroMax ET installed.
When you open the
SynchroMax
dialog from
Instrument >
SynchroMax Automation
, ScreenWorks Software automatically goes
into
Remote Mode
, giving control of the instrument to the
SynchroMax ET software. ScreenWorks Software returns to
Manual
Mode
when the
SynchroMax
dialog closes.
To run a series of experiments with the SynchroMax ET, first open the
protocol file that you want to use in ScreenWorks Software. Only one
protocol is used in a run, as many times as the SynchroMax ET supplies
plates for it.
Having prepared the protocol, open the
SynchroMax
dialog and select
a
Stack Layout Template
. These files specify, for each plate location
in the five-position stage, how many plates will be loaded to that
position, from which stack, and where they will be delivered to after
use. A number of
Stack Layout Templates
are supplied—select the
one that suits your assay.
When a
Stack Layout Template
is opened in the dialog the graphic in
the dialog displays which stacks plates should be loaded in. You can use
this as a guide for loading the plates. All plates in a single stack should
be of the same type.
If there are any stage positions that will use the same plate throughout
the duration of the run, these should be loaded manually before the run
begins. Stage positions that will have plates brought to them during the
run should be empty at the start of the run.
Save
When clicked, saves the signal test as an ASCII text file
(*.sig).
When clicked, prints the signal test.
Image Viewer
Displays the CCD image showing where RLU values were
derived for the signal test.
Show > Hide Mask
—Shows or hides the mask used in
the test, indicating the pixel area used to derive the RLU
value for each well.
View Image
—Refreshes signal test image.
Item
Description
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...