Troubleshooting
220
0112-0109 H
222: “Unable to unload plate.
Plate already detected at landing
pad.”
Plate already present
on landing pad.
Remove plate on
landing pad prior to
unloading a new
plate.
223: “Unable to unload plate. No
plate detected at requested
position.”
No plate present in
position.
Clear error and
proceed with
experiment.
Plate may have
popped out of
position.
Clear plate that is
out of position prior
to resuming
experiment.
Bar code reader misreads the
bar code and file name reads
“Bad_bar_code”.
Misplacement of bar
code or incorrect
type of bar code.
Bar code must be on
the side of the plate
where the bar code
is reading. Place
label as low as
possible. Use labels
with numbers on top.
Table 9-10
Troubleshooting Robotic Integration (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...