FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
151
3.
Select the appropriate
Plate Format
from the 96, 384, or 1536
options and select the
LED/Filter
pair you want to calibrate
from the drop-down list.
4.
Place the
Flat Field Calibration Plate
into the
Read Position
.
The
Flat Field Calibration Plate
may be a Molecular Devices
supplied plate or a
Read Plate
loaded with an appropriate dye
for the LED/Filter wavelength pair being calibrated.
Molecular Devices supplied yellow, 96-, 384- or 1536-well plate
can be used to
Flat Field
calibrate the following LED/Filter
pairs.
390–420 nm/565–625 nm
420–455 nm/475–505 nm
420–455 nm/515–575 nm
470–495 nm/515–575 nm
470–495 nm/565–625 nm
495–505 nm/526–586 nm
510–545 nm/565–625 nm
610–626 nm/646–706 nm
The following dye should be used to calibrate the respective
LED/Filter pair.
360–380 nm/400–460 nm with MQAE (detailed instructions
for handling MQAE are outlined in the MQAE section).
390–420 nm/440–480 nm with Coumarin (detailed
instructions for handling Coumarin are outlined in the
Coumarin section).
5.
Once the appropriate plate is placed in the
Read Plate
position,
select the
Flat Field Calibration
button. The calibration is
complete when the
Assay Finished (Unlock)
light is
illuminated on the instrument and the LED/Filter pair is
identified as
Done
under the
Flat Field
cal column.
6.
When complete, select
Close
to exit the
Calibration
dialog.
7.
Place your yellow signal test plate of the appropriate well format
into your system.
8.
Perform a signal test by selecting
Yellow Plate Signal Test
from the
Instrument/Manual Operation
menu, or using the
corresponding button. The yellow plate standard deviation
should be 5% or less.
Note:
None indicates either no LEDs are illuminated or the
instrument used a blank emission filter position to create the
calibration file.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...