Troubleshooting
198
0112-0109 H
Troubleshooting Start-Up
Table 9-2
Troubleshooting Start-up
Symptom or Error
Message
Possible Causes
Solutions
“Connecting to the
instrument, please wait. Will
enter offline if the
instrument is not connected
or if this dialog is closed.
Dialog may not close
immediately on pressing the
close button while
connection is being
attempted.”
ScreenWorks
Software is waiting
for the FLIPR
®
System to complete
initialization prior to
connecting.
Wait until
Assay Finished
(Unlocked)
light is the
only light illuminated on
the front instrument panel
before the software and
instrument will connect.
"Camera chiller subsystem
not functional. Please
contact Technical Support.
The instrument may have
degraded read performance
until it is repaired: Can't
communicate with chiller."
Instrument is not
communicating with
the chiller.
Make sure the chiller is
turned ON.
Contact Technical Support
at
+1-800-635-5577.
409:“Initialization halted
because tips are loaded.
Please verify that the
appropriate tip rack
(96,384,1536) is loaded and
select RESET”.
Tips are on the
pipettor head and
need to be
unloaded during
instrument
initialization/reset.
Place tip rack in source
3/tip loading position to
unload tips and select
Reset
from the
Instrument
menu.
410: “Initialization halted.
Either the upper or lower
door is open. Please shut
both doors and select
RESET”.
The door is open
during instrument
initialization/reset.
Close the upper and lower
doors and select
Reset
from the
Instrument
menu.
411: “Front Panel Interrupt
button was pushed while
initializing. Initializing was
not completed. Select Reset
when ready to continue.”
The front panel
interrupt button
was pushed,
stopping instrument
initialization.
Select
Reset
from the
Instrument
menu.
“Disconnected” is displayed
in the bottom of the
Instrument Status
Window
.
ScreenWorks
Software is in
Offline Mode
.
Select
Go Online
from the
Instrument
drop-down
menu.
Multiple software
copies are open.
Close all versions of
ScreenWorks Software
before starting software.
Instrument
timed-out.
Select
Go Online
from the
Instrument
drop-down
menu.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...