Troubleshooting
214
0112-0109 H
“Number of reads before
dispense are too large.”
Number of reads prior
to dispense are larger
than the number of
reads in the first
interval. See appended
string on the error
message for details.
Decrease the number of
reads before dispense so
it is within the specified
range.
238: “Protocol requests
greater than 800 reads.
This is not permitted.”
Sum of all reads for all
read actions and
configurations must not
exceed 800.
Reduce the number of
reads in the experiment
to less than 800 for all
combined read modes.
239: “Protocol requests
greater than 50 images.
This is not permitted.”
Sum of all raw images
for all read actions and
configurations must not
exceed 50.
Reduce the number of
raw images in the
experiment to less than
50 for all combined
dispenses.
400: “Camera is not
functional. Please contact
Technical Support. You
will not be able to run
protocols until it is
repaired.”
Camera is not
functional.
Contact Technical Support
at
+1-800-635-5577.
Table 9-5
Troubleshooting the Optics (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...