Robotic Integration
266
0112-0109 H
Example Command Line
washcellreservoir<TAB>fluid 2, 6, waste a, 8, 1, 0, high<CR>
Example Response
Ok<CR>
Suggested Steps in Command usage
•
Make sure an experiment is not running.
•
Ensure that tips are loaded.
•
Send
washcellreservoir
command.
•
Response will be OK, confirming that command execution has
begun, or a Condition code related to the command indicating
that execution could not begin.
•
Monitor status until the wash is complete.
Command Specific Error Responses
C77–Parameter is out of range. The range is ...
C78–The number of parameters for wash
Cell Reservoir
commands is
invalid.
C79–Invalid volume parameter. It must be 'low' or 'high'.
C72–Hold time is too small.
C73–Hold time is too large.
Hold time
float
Pause time in seconds,
leaving the fluid in the
reservoir for desired
period of time before
draining it.
0–300
Volume
string
This indicate the level of
fluid to be filled in the
Cell Reservoir.
low, high
Table A-2
Washcellreservoir<TAB>Fluid Type, Fill Speed, Drain
Destination, Drain Speed, Wash Cycles, Hold Time, Volume Level<CR>
command parameters (cont’d)
Parameter
Type
Description
Acceptable values
or range
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...