Troubleshooting the Pipettor
Table 9-4
Troubleshooting the Pipettor
Symptom or Error
Message
Possible Causes
Solutions
Cannot uninstall pipettor
head.
The D-Axis Knurled
Nut is caught on the
D-Axis receiver.
Grasp the D-Axis Knurled
Nut receiver and press
downward to give the
pipettor head additional
clearance from the
D-Axis Knurled Nut.
Pipettor head is
caught on the
alignment pins.
Lift the pipettor head up
prior to moving left to lift
the pipettor head off of the
pipettor mount alignment
pins.
Cannot install pipettor
head.
The Head Clamp
Knob is not
tightening all the
way down.
Slide the pipettor head all
the way to the right over
the pipettor mount. When
installed properly on the
alignment pins, the pipettor
head should not slide left
without lifting.
D-Axis Knurled Nut
does not reach
receiver threads.
Lift D-axis receiver to
engage threads and then
tighten the
D-Axis Knurled Nut.
231: “Pipettor head not
detected. Please verify
proper installation and
select
Reset
.”
No pipettor head is
present on pipettor
mount.
Install appropriate pipettor
onto instrument.
Pipettor head is not
installed properly.
Head clamp knob may not
be secured tightly for
instrument to identify
pipettor head. Reinstall
pipettor head.
The open protocol
may not use the
pipettor on the
instrument.
Install appropriate pipettor
onto instrument.
Change plate format used
in the software to match
the pipettor head.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...