FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
219
Troubleshooting Robotic Integration
Data parameters need to
be reset (subtract bias,
spatial uniformity
correction, positive
control,
etc
.) for each
experiment.
*.vamp file
parameters not set up
correctly.
Select the
Grouping
or
Correction
buttons to
adjust the parameters as
necessary.
False negative wells.
Pipettor adjustment
problem.
Tip problem.
Contact Technical Support
at
+1-800-635-5577.
Table 9-9
Troubleshooting Data (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Table 9-10
Troubleshooting Robotic Integration
Symptom or Error
Message
Possible Causes
Solutions
220: “Unable to load plate. No
plate on landing pad.”
Plate present on
plate landing pad,
but plate presence
was not detected.
Contact Technical
Support at
+1-800-635-5577.
If plate not
positioned properly
by robot, contact
Technical Support for
robot manufacturer.
No plate was
delivered to landing
pad by 3
rd
party
robot.
Contact Technical
Support for robot
manufacturer.
221: “Unable to load plate. Plate
already detected at requested
location.”
Plate already at
requested position.
Unload plate
currently in the
desired position prior
to delivering new
plate.
If robot is attempting
to load multiple
plates to one position
without retrieving
plates, contact
Technical Support for
robot manufacturer.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...