Robotic Integration
262
0112-0109 H
bar_code<SP>installed<TAB>
plate_handler<SP>installed<TAB>
cell_reservoir<SP>installed<TAB>
camera_installed<SP>emccd<TAB>
chiller_installed<SP>installed<TAB>
wave1<SP>470-495 nm<TAB>
wave2<SP>510-545 nm<TAB>
emfilt1<SP>515-575 nm<TAB>
emfilt2<SP>515-575 nm<TAB>
emfilt3<SP>515-575 nm<CR>
Suggested Steps in Command Usage
•
Send
configuration
command.
•
Response will be instrument configuration.
Statusex<CR>
Command Description
The
statusex
command is an extension of status command. It returns
the status of the instrument in more details than the status command.
The
statusex
response will be an ASCII text string which includes all
status command’s response information plus:
•
Tip washer status
•
Cell Reservoir status
•
Camera chiller status
•
Camera intensifier status
Example Command line
statusex<CR>
Response format
instrument <SP>instrument name<TAB>
version<SP>mm/dd/yyyy<TAB>
function<SP>ok|errorcode|fatalcode,[errorstring]<TAB>
inst_status<SP>busy|fault|error|resetting|abort|idle,remote|manual,o
ffline|online<TAB>
exp_running<SP>yes|no<TAB>
readplate<SP>platename|none,have|empty<TAB>
sourceplate1<SP>platename|none,have|empty<TAB>
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...