Data Processing Algorithms
280
0112-0109 H
The results show only numbers above zero. The subtract bias algorithm
places the data in a readily interpretable form. The positive control
wells at Sample 4 range from 93–109% positive. In contrast, the
negative control wells remain close to zero percent positive throughout
the experiment. The experimental wells show two wells with about 90%
activity and one with 79% activity at Sample 4.
Response Over Baseline
The response over baseline correction algorithm compensates, to a
certain extent, for variations within a single well.
Determining Response Over Baseline Correction
The correction factor is derived by calculating the mean fluorescence
counts in each well between the baseline start and end samples (see
below).
All samples taken from a particular well are divided by their well-
specific mean baseline correction factor. For example, all samples from
A1 are divided by 8200, A2 by 8533,
etc
.
The results of applying the response over baseline correction factor are
presented in the table below. Note the data is displayed is displayed as
Sample
Time
Well
A1
– Ctrl
A2
– Ctrl
A3
– Ctrl
A4
Exp
A5
Exp
A6
Exp
A7
+ Ctrl
A8
+ Ctrl
A9
+ Ctrl
1
0
8000
8500
9500
8200
9200
8800
10000 9500
8700
2
5
8200
8600
9400
8300
9000
9000
9500
9700
9000
3
10
8400
8500
9200
8500
9000
9100
9700
9700
9400
Mean
8200
8533
9367
8333
9067
8967
9733
9633
9033
Note:
In actual practice, data from all 96 wells are included in the
calculations, but in this example, data from only nine wells are
presented, to simplify the examples.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...