Troubleshooting
204
0112-0109 H
408: “Plate format file
fault. Please contact
Technical Support. You will
be unable to use this plate
until repaired.”
Plate format file fault. Select
Reset
from the
Instrument
menu.
Contact Technical Support
at
+1-800-635-5577.
ScreenWorks indicates the
user-defined name of the
data file is invalid.
Illegal characters
were used in the file
name.
Eliminate illegal
characters from user-
defined name (for
example
,
“
, ’ : ? ; )
.
User-defined name
includes too many
characters.
Use 25 characters or less
in the user-defined name.
Saturation detected.
One or more wells in
the microplate
saturated the camera.
A saturation warning in
either the signal test
dialog or on a plate will
prevent you from reading
the plate. It is at the
discretion of the user to
decide whether to address
the saturation warning or
proceed forward. If
multiple wells within a
plate are saturating the
camera, decrease the
Excitation Intensity,
Exposure Time
, or
Camera Gain
to prevent
saturation.
Table 9-3
General Troubleshooting (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...