Running an Experiment
160
0112-0109 H
components are installed, see the appropriate section in
for installation instructions.
Checking the System
Once a day, before running your first assay plate, run a
Yellow Plate
Signal Test
to ensure the system is performing according to
specifications. A yellow signal test plate is available for each
corresponding pipettor head supplied with the system.
CAUTION!
Store the yellow signal test plate in a safe place away from
bright light on an even surface. Avoid scratching the plate bottom as
scratches can affect the standard deviation. A signal test plate does not
need to be run more than once a day unless system components such
as pipettor heads or LEDs are changed.
Running the Yellow Plate Signal Test
To run a Yellow Plate Signal Test:
1.
Select
Yellow Plate Signal Test
from the
Instrument >
Manual Operation
menu or click the
Yellow Plate signal test
button.
2.
Set the following parameters:
3.
Select the correct plate from the
Select Plate
list.
4.
Select appropriate
Read Mode
.
5.
Select the appropriate
Excitation/Emission Wavelengths
.
6.
Set the
Excitation Intensity
to 50.
7.
Set the
Exposure Time
to 0.1 seconds.
Note:
Please do not attempt to change pipettor heads, wash
reservoirs, LED banks and filters while the instrument is running. All
features must be exchanged in the appropriate instrument mode.
Optics calibration will need to be performed when the instrument is
started for the first time after re-configuration of LEDs or emission
filters.
Note:
Make sure that the system is turned on and the CCD camera is
cooled to -60 °C ± 2 °C when the EMCCD camera is present or -
20 °C ± 5 °C when the ICCD camera is installed to ensure proper
system functioning.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...