Data Processing Algorithms
278
0112-0109 H
Positive Control Scaling
The positive control algorithm compares the percent change in
fluorescence counts of the positive control wells with all wells. This
algorithm facilitates comparisons of results between data runs (
i.e.,
different plates) by controlling factors such as cell density, cell
response, laser power, or exposure time. This algorithm also makes
EC
50
comparisons easier.
Determining the Positive Control Scaling
1.
The means of all samples of the positive control wells are
calculated.
2.
The difference in fluorescence counts between Sample 1 and all
of the samples is calculated.
3.
The greatest difference is determined.
4.
100 is divided by the greatest difference in fluorescence counts
(determined above) to give the positive control correction factor
e.g.,
100/42296 = 0.0024.
5.
All samples are multiplied by the correction factor 0.0024 (see
the following table).
Sample
Time
Well
Mean
Difference
from
Sample 1
A7 + Ctrl A8 + Ctrl A9 + Ctrl
1
0
8900
8930
8961
8930
0
2
5
8370
9027
9177
8858
-72
3
10
8547
9027
9585
9053
123
4
20
50223
48391
55064
51226
42296
5
25
45755
45590
49955
47100
38170
6
30
43165
46502
49955
46541
37610
Sample Time
Well
A1
– Ctrl
A2
– Ctrl
A3
– Ctrl
A4
Exp
A5
Exp
A6
Exp
A7
+ Ctrl
A8
+ Ctrl
A9
+ Ctrl
1
0
21
21
21
21
21
21
21
21
21
2
5
21
21
21
21
20
21
20
21
23
3
10
22
21
20
22
20
22
20
21
23
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...