Precipitation in the Reagent Buffer
The FLIPR
®
Calcium Assay Kit is compatible with numerous buffers. Use
buffers shown to work in previously established assays, if available.
Response is Smaller than Expected
The agonist and antagonist may stick to the tips and trays. Use up to
1% BSA in all compound buffer diluents and presoak tips in compound
buffer with up to 1% BSA.
Apparent Well-to-Well Variation is Observed
AquaMax DW4 or equivalent dispenser is recommended for use with
additions of cells or dye prior to the assay on the FLIPR
®
Tetra System if
apparent well-to-well variation is observed. In some cases allowing the
plates to stand at room temperature prior to use or adding a single mix
cycle in the compound or assay plate may decrease well-to-well
variation.
FLIPR Membrane Potential Assay Kit Protocol
This section provides the following information and procedures:
•
Materials required for the assay
•
Cell preparation guidelines
•
Dye loading procedure
•
Troubleshooting tips for membrane potential assays
Required Materials
The following materials are required for running a membrane potential
assay:
Note:
Do not use the same source plate for presoaking and compound
addition when using a 384 Pipettor head in the FLIPR
®
Tetra System.
Instead, use a 'Boat' for the presoak.
Item
Source
FLIPR
®
Tetra System with Membrane Potential
Optics Kit installed:
FLIPR
®
Tetra System LED Module 510–545 nm
FLIPR
®
Tetra System Emission Filter 565–625
nm
Molecular Devices
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...