FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
91
3.
To assign a single group concentration, click the
Group
tab and
do the following:
Enter a name for the group (for example
,
compound name)
in the
Group Name
field.
Select or enter the concentration in the
Concentration
box.
Select the concentration units in the
Units
drop-down list.
The choices are nM,
μ
M, mM, M or Log.
In the
Notes
area, type in any notes that you want
associated with the specified group and data file.
4.
To assign a series of concentrations, click the
Series
tab and do
the following:
Enter a name for the group (for example
,
compound name)
in the
Group Name
field.
Select or enter the lowest or highest concentration of the
series in the
Starting Value
field.
Select or enter the change in concentration between
concentrations in the series in the
Step Increment
field.
Select the concentration units in the
Units
drop-down list.
The choices are nM,
μ
M, mM or M.
Decide how the
Step Increment
is applied to the
Starting
Value
to create the series, for example, Plus, Minus,
Multiply or Divide, and select it from the
Operation
drop-
down list.
If you have replicates, select whether they are aligned in row
or column format on the plate. If replicates align in both row
and column formats, indicate the number of replicates (for
example
,
Replicates
in rows indicates all wells in the
selected rows will have same concentration). If two rows are
to include the same concentration, indicate “2” replicates. If
there are no replicates, select
None
.
Select the direction in which the series will increment, for
example, Left, Right, Top or Bottom.
In the Notes area, type in any notes that you want
associated with the specified group and data file.
5.
When finished, click
OK
to save the defined group.
When a series is created, assign wells to it as described above.
Successive rows or columns are assigned different concentrations and
group color automatically, as configured in the series.
Note:
The
Series
option is typically used for assigning a
dose-response curve with regular increment steps.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...