Robotic Integration
248
0112-0109 H
removeplates<SP>none<TAB>
platehandler<SP>idle<TAB>
plate_on<SP>no<TAB>
camera_temp<SP>ok,-35,-35<TAB>
chamber_temp<SP>off,0,0<TAB>
waste_bottlea<SP>ok<TAB>
waste_bottleb<SP>ok<TAB>
fill_bottlea<SP>ok<TAB>
fill_bottleb<SP>ok<TAB>
tips<SP>on<TAB>
outer_auto_door<SP>closed<CR>
Suggested Steps in Command Usage
•
Send status command
•
Response will be instrument status
Loadplate<TAB>Location, Last Plate[, BAR CODE]
<CR>
Command Description
Loads plate or tips to the specified location from landing pad. The
plates or tips must have been placed on the instrument landing pad
prior to the command being issued. Up to 13 plates (1 source plate, 12
compound plates) can be used in single experiment.
Additional parameters are available to provide the bar code for the
delivered plate and identify the final plate for an assay.
The bar codes of up to 5 plates can be recorded in the data file.
Command Parameters
There are four parameters available for use with this command.
The parameters are location, bar code, and last plate. The location
parameter is required. The bar code and last plate parameters are
optional. If the bar code parameter is used then the plate order
parameter must also be used.
Location parameter must be one of following:
•
re–which indicates read plate location.
•
s1–which indicates source plate 1 location.
•
s2–which indicates source plate 2 location.
•
s3–which indicates source plate 3 location.
•
ti–which indicates tips loading location.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...